Spectrophotometer Model Genesys 50

CODE : 3019205

CALL

Last Update : 04/06/2025

Availability : สินค้า by order ( รอเข้าคลังประมาณ - วัน )

Features

  1. It is a spectrophotometer that measures the absorbance of samples using ultraviolet and visible light ranges.
  2. The optical system is Dual Beam.
  3. It has a Spectral Bandwidth of 2 nm.
  4. The light source is a Xenon lamp.
  5. It has a Dual Silicon Photodiodes detector system.
  6. Wavelength can be continuously selected for use in the wavelength range from 190 to 1100 nm.
  7. Wavelength Accuracy is not more than ± 0.5 nm.
  8. Wavelength Repeatability is ± 0.2 nm.
  9. It can measure absorbance in the range of -2A to 3.5 A.
  10. It can display photometric values in the range of -3A to 5A.
  11. Photometric Repeatability is ± 0.001A at 1A.
  12. Noise is less than or equal to 0.00020 A at 0 A, 0.00030 A at 1 A, and 0.00040 A at 2 A at wavelengths of 260 nm and 500 nm.
  13. Drift is not more than 0.0005 absorbance units per hour.
  14. Stray light is not more than 0.05%T at 220 nm and 0.03%T at 340 nm.
  15. The color display is a 7-inch touch screen, capable of displaying numbers and graphs.
  16. It comes with a sample holder that can hold 1 cuvette.
  17. It has direct-use programs with the following analysis capabilities:
    1. Measures Absorbance, Transmittance, and concentration of samples.
    2. Determines the concentration of samples compared to a standard curve. Can create standard curves.
    3. Can perform continuous scanning throughout the wavelength range from 190 to 1100 nm (Scanning).
      • Maximum scanning speed: 1,600 nm per minute.
      • Data resolution can be selected as: 0.2, 0.5, 1, 2, and 5 nm.
    4. Measures kinetic reaction rates (Kinetics).
    5. Includes Performance Verification Tests program to check the accuracy of the instrument.
  18. It has USB ports for connecting to a computer, keyboard, or mouse.
  19. Data can be stored using a Flash memory device via a USB port.
  20. Includes the following accessories:
    • 1 dust cover.
  21. Operates on 220 Volts, 50 Hertz.