Performance Data Temperature
Temperature range 5 °C above ambient temperature to [°C] : 300
Temperature variation at 150 °C [± K] : 2.3
Temperature fluctuation at 150 °C [± K] : 0.4
Heating-up time to 150 °C [min] : 50
Recovery time after 30 seconds door open at 150 °C [min] : 19
Interior volume : 57L
Doors and Shelves :
Unit doors : 1
Shelves : 2
Internal Dimensions (W×H×D) : 360x420x380 mm
External Dimensions (W×H×D) : 560x625x565 mm
Electrical data : 230V , 50/60 Hz